Can You Have an MRI With Metal in Your Leg?

For most people with metal in the leg, an MRI is possible and safe. That includes joint replacements, plates, screws and rods. For most, the bigger practical issue is not danger but getting clear pictures and giving the scanning team the right device details. A few kinds of metal, mainly frames that sit outside the skin and fragments from gunshots or blasts, need a closer look before scanning.1mrisafety.com. Can You Have an MRI With Metal in Your Leg?

Why Implants Fixed in Bone Are Low Risk

Titanium, common in modern implants, is only very weakly affected by an MRI magnet.2PubMed Central. Are titanium implants actually safe for magnetic resonance imaging examinations? Hip replacements tested in a scanner magnet showed pull and twisting forces too small to measure.3PubMed. The effect on medical metal implants by magnetic fields of magnetic resonance imaging Over about 30 years of testing, nearly all studies found non-magnetic implants safe to scan.4PubMed Central. Are titanium implants actually safe for magnetic resonance imaging examinations? Titanium and most cobalt-chromium alloys are weakly or non-magnetic, while some older stainless steels are more magnetic. That is why a surgeon’s operative note listing the maker and model tells you more than a guess based on how old the hardware is.

Even strongly magnetic hardware stays put when bone holds it. One highly magnetic spinal implant felt pulling forces many times higher than any orthopedic implant recorded before. Because it was rigidly fixed in bone, movement was judged highly unlikely.5PubMed. Evaluation of the translational and rotational forces acting on a highly ferromagnetic orthopedic spinal implant in magnetic resonance imaging Implants firmly fixed to bone are not displaced by MRI, and the same principle applies to leg hardware.6PubMed Central. Are titanium implants actually safe for magnetic resonance imaging examinations?

Joint replacements aren’t yanked around in the scanner either. Hip and knee components are large but mostly weakly magnetic, so the force on them is tiny next to what holds them, which is bone, cement or bone growth.

Heating is the theoretical worry. The scanner’s radio waves can, in theory, warm metal by inducing small electric currents in it.7PubMed Central. Are titanium implants actually safe for magnetic resonance imaging examinations? Yet no confirmed reports exist of tissue damage from MRI heating near orthopedic implants.8PubMed. Key safety considerations for MRI examination of patients with an orthopaedic implant. In a group of patients with titanium bone-anchored limb implants, no one had pain, heat or loosening after MRI.9PubMed Central. Clinical implementation of MRI for evaluation of titanium press-fit osseointegration implants Heating depends on shape. Long, thin or looped metal is a bigger concern than a compact block. Knee and hip replacements form no long loops, and screws and plates are short and sit inside bone, so they have little chance of acting as an antenna.

Feeling warm during the scan doesn’t mean the implant is heating; the scanner itself commonly causes warmth, especially in stronger magnets. Still, tell the technologist through the intercom if any spot feels hot or painful, and they can stop.

Scanners come in strengths such as 1.5 and 3 tesla, and many implants have been tested at 3 tesla for pull and twist. Your device’s label may name which strength it was tested at.

Metal That Needs Extra Care

An external fixator is a frame of pins and bars outside the skin holding a broken bone. Its length and the loops its parts form can carry current and heat up in the scanner.10MRIsafety.com. Can You Have an MRI With Metal in Your Leg? The heating concern is mostly at the pin sites, where metal meets skin and soft tissue, which is why testers measure those spots.

Since 2014 these external frames have been labeled “MR Conditional” rather than “MRI compatible,” because they can be assembled in endless configurations.11PubMed Central. Safety and utility of magnetic resonance imaging of patients with external fixators: a systematic review In cadaver testing at 1.5 tesla, pin sites warmed about 0.3°C, and only one test passed the 2°C level regulators treat as a concern.12PubMed Central. Safety and utility of magnetic resonance imaging of patients with external fixators: a systematic review In a cadaver test of leg traction pins, the pins didn’t heat beyond room temperature, move or loosen.13PubMed. A cadaveric simulation of distal femoral traction shows safety in magnetic resonance imaging If you wear a frame, the team needs its brand and parts, since safety depends on the exact setup. Scan settings and the frame’s position in the scanner affect heating, so expect them to plan conditions in advance rather than scan on the spot.

Bullets and shrapnel depend on what they’re made of and where they are. Most bullets and pellets tested were not magnetic, but many contain traces of magnetic metal.14mrisafety.com. Can You Have an MRI With Metal in Your Leg? Magnetic ammunition tended to be foreign-made or military.15mrisafety.com. Can You Have an MRI With Metal in Your Leg? Shrapnel usually contains steel, so it’s treated as a potential hazard.16mrisafety.com. Can You Have an MRI With Metal in Your Leg?

Location matters because the risk is that a magnetic fragment could shift or twist. Fragments in bone, tendon or muscle are much less likely to move enough to cause harm than those near vital structures.17Questions and Answers in MRI. MR safety bullets Age counts too. Fragments in place for months or years are usually held by scar tissue, and one review suggests waiting at least 6 weeks before scanning a likely magnetic fragment in a non-critical spot.18Questions and Answers in MRI. MR safety bullets

For a retained bullet, the team may review old X-rays or CT scans to see where the fragment sits. They often assume it is magnetic unless shown otherwise, and they weigh how much the scan will help against the risk.

Vein filters and stents are less worrying. People with vascular stents have been scanned, including at 3 tesla, with no adverse event ever reported when label instructions are followed.19ISMRM. Guidelines for the Management of Patients with Vascular Stents Referred for MRI Examinations Filters placed in the large abdominal vein (IVC filters) have not caused complications or displacement on MRI.20Circulation. ACC/AHA 2007 Guidelines for the Management of Patients With Peripheral Arterial Disease For a few weakly magnetic steel filters, waiting at least 6 weeks is advised unless the scan is urgent.21Circulation. ACC/AHA 2007 Guidelines for the Management of Patients With Peripheral Arterial Disease

What MR Safe, MR Conditional and MR Unsafe Mean

A 2005 standard set three labels for devices:

Conditional isn’t a warning. Almost any metal device will be conditional, since only non-metal items can be considered to have no known MRI risk. The conditions might list a maximum magnet strength or how much energy the scan may use, and the technologist sets the scanner to fit them. Older cards may say “MRI compatible,” an outdated term. Because the label belongs to the device and its stated conditions, a device that is safe under set conditions at one magnet strength may not be listed for another.

You won’t see these labels on the device itself, since it is inside you. They appear in the maker’s instructions, on implant cards and in the hospital’s device database. Staff look up your implant’s entry and match it to the scanner they plan to use.

What to Tell the MRI Team Before Your Scan

Before scanning, staff must confirm the implant’s type, location, make and model, including any wires or attached parts.25Image-wisely. Basics of MRI Patient Screening They then look up its safety information from the maker.26Image-wisely. Basics of MRI Patient Screening That can come from surgical notes, a device ID card, doctors’ notes or a past MRI screening form.27Image-wisely. Basics of MRI Patient Screening

To track down details, call the hospital where surgery took place and ask for the operative note, and check wallet cards or discharge paperwork for the brand and model. Give the information days ahead. If that hospital has closed or merged, ask the records department of the system that took it over. Your current surgeon or primary care office may have older letters or clinic notes that mention the implant. A past X-ray or CT can also show its shape and position, and staff may use it to help decide how to proceed.

Before any MRI you fill out a safety questionnaire that asks about surgeries, implants, injuries involving metal and sometimes jobs like welding. Answer it fully, including things you think are minor, because staff use your answers to decide whether more checks are needed. The technologist reviews the form again in person, so you can clarify anything there.

Not knowing exactly what you have doesn’t automatically rule out a scan. Many people have metal that can’t be fully identified.28University of Wisconsin–Madison. Guidelines for MRI Clearance of Patients with Metallic Implants of Uncertain Identity Guidelines recognize the risk is often very low and outweighed by what the scan shows, so scanning can still go ahead.29University of Wisconsin–Madison. Guidelines for MRI Clearance of Patients with Metallic Implants of Uncertain Identity Hospitals may use a formal, signed-off procedure for scanning such patients.30PubMed Central. A framework for developing generic implant safety procedures for scanning patients with medical implants and devices in MRI

Timing after surgery is rarely an obstacle. Weakly magnetic hardware rigidly fixed in bone, such as screws, can be scanned right after surgery, and no healing wait is generally needed for these.

Will the Metal Spoil the Pictures

Implants can create distortions that hide nearby tissue from the reading doctor.31mrimaster. MARS MRI | metal artifact reduction mri Metal disturbs the magnetic field around it, so signal from nearby tissue gets bent and lost, producing dark or bright patches. The effect reaches only a limited distance, so scans of areas far from the implant are usually clearer. How large the blurred area is depends on the metal type, its size and how it sits in the scanner. Titanium typically causes less distortion than stainless steel or cobalt-chromium, and a long rod or full joint replacement distorts more than a single small screw.

MARS MRI is a set of scan techniques designed to cut metal blur and show soft tissue near implants.32PubMed Central. The Role of Advanced Metal Artifact Reduction MRI in the Diagnosis of Periprosthetic Joint Infection These metal-reducing scans can reveal problems around joint replacements that used to stay hidden.33PubMed. MR Imaging with Metal-suppression Sequences for Evaluation of Total Joint Arthroplasty They can mean longer scans or grainier images, though newer methods keep scan time practical.34PubMed. Advances in MRI around metal In testing with hip implants, blur was less on 0.55- and 1.5-tesla scanners than 3-tesla.35PubMed Central. Visual and quantitative assessment of hip implant-related metal artifacts at low field MRI: a phantom study comparing a 0.55-T system with 1.5-T and 3-T systems You can ask whether a metal-reducing scan or a lower-strength scanner is planned.

For a painful joint replacement, doctors often need to see the soft tissue around it to check for infection. Metal-reducing scans can show fluid buildup, swelling and a thickened joint lining linked to infection. In one group, finding no joint fluid made infection very unlikely, though that is one clue among several. That is why your doctor may order an MRI even with an implant.

Alternatives When MRI Isn’t Possible

Dual-energy CT can show bone bruising in the knee about as well as MRI and handles metal implants well.36PubMed. Dual-energy CT for detection of knee fractures as alternative to MRI It can reduce metal blur and help diagnose hardware problems.37PubMed. Dual energy CT in musculoskeletal applications beyond crystal imaging: bone marrow maps and metal artifact reduction For bone infection in the leg, it found about 89% of cases, as many as MRI, but gave more false alarms.38PubMed Central. Osteomyelitis of the Lower Limb: Diagnostic Accuracy of Dual-Energy CT versus MRI MRI is still the best overall test for knee injuries.39PubMed. Dual-energy CT for detection of knee fractures as alternative to MRI

Dual-energy CT takes images at two different X-ray energy levels, which lets software separate materials such as bone, metal and soft tissue. That separation helps reduce the streaks metal causes on a regular CT. It is done in a CT scanner like the one used for ordinary CT, so the experience is much the same. It can answer certain questions, but MRI shows ligaments, cartilage and soft tissue in more detail, so some questions still need it. Whether it fits depends on what your doctor is trying to find out.

If your MRI is declined, ask exactly which device or fragment caused the concern. Ask whether a different scanner strength would allow it, and whether dual-energy CT can answer the question. Remember that CT uses X-ray radiation, unlike MRI.