A Computed Tomography (CT) scan provides cross-sectional images of the body’s internal structures. Unlike standard X-rays, which compress three-dimensional objects into a two-dimensional image, CT scans use complex calculations to reconstruct detailed, slice-by-slice views. The foundational measurement allowing CT scans to differentiate between tissues like bone, fat, and muscle is the Hounsfield Unit (HU) scale. This standardized numerical scale quantifies the density of materials within the body, making subtle differences in tissue composition visible.
What Hounsfield Units Measure
Hounsfield Units are a quantitative measurement of radiodensity, reflecting how much a specific tissue absorbs or “attenuates” the X-ray beam passing through it. The physical density of a material is directly proportional to its X-ray attenuation. The HU scale translates this X-ray absorption into a numerical value consistent across different CT scanners.
The scale is a linear transformation based on two reference substances: water and air. Distilled water is arbitrarily defined as zero Hounsfield Units (0 HU), serving as the mathematical reference point. Air is assigned a value of -1000 HU because it is the least dense material typically encountered in the body and attenuates virtually no X-rays.
Materials denser than water absorb more radiation and are assigned positive HU values. Conversely, materials less dense than water absorb less and receive negative HU values. This standardization allows radiologists worldwide to communicate precisely about tissue composition simply by referencing its HU value.
Reading the Hounsfield Scale
The Hounsfield scale spans a wide range, typically from -1000 HU to over +3000 HU, with each segment corresponding to a different tissue type. Understanding these ranges is crucial for interpreting CT images and differentiating between tissues. The lowest end is air, registering at approximately -1000 HU, which makes structures like the lungs or sinuses appear black on the scan.
Fat tissue is less dense than water and falls within a negative range, typically between -50 HU and -100 HU. Soft tissues, such as muscle, liver, and brain matter, are closer to the density of water and generally occupy the range of +20 HU to +70 HU. For example, muscle is typically between +10 HU and +40 HU.
The highest end of the scale is reserved for calcified or mineralized structures that strongly attenuate the X-ray beam. Dense bone, or cortical bone, registers values from +400 HU up to +1000 HU or more, appearing bright white on the image. Even denser materials, such as metal implants, can exceed +3000 HU.
Diagnostic Importance of Hounsfield Values
Quantifying tissue density using Hounsfield Units provides objective data fundamental to clinical decision-making. Precise HU measurements characterize lesions, such as adrenal masses. An adrenal mass measuring less than 10 HU on a non-contrast scan is often confirmed to be a benign, lipid-rich adenoma. This low HU value indicates high fat content, effectively ruling out many malignant possibilities.
HU values are also indispensable in evaluating bleeding, as the density of blood changes predictably over time. Acute hemorrhage (recent bleeding) has a high protein concentration and often measures between +50 HU and +90 HU. In contrast, older, chronic fluid collections or subacute hematomas have lower HU values, typically ranging from +25 HU to +35 HU, allowing clinicians to accurately estimate the age of a hemorrhage.
Kidney Stone Analysis
In urology, the HU measurement of kidney stones is a predictive tool used to guide treatment planning. Stones with lower HU values, such as those composed of uric acid (often less than 450 HU), are generally softer and more likely to be successfully broken apart by non-invasive shock wave lithotripsy (ESWL). Conversely, stones with high HU values, such as calcium oxalate or cystine stones, are denser and may require more aggressive surgical interventions.
Quality Control
Hounsfield Units also serve as a quality control measure. Technicians regularly check that a vial of water in the scanner registers near 0 HU, ensuring the equipment is correctly calibrated for all diagnostic procedures.

